Recall Z-1151-2018 ยท JQP Calculator/Data Processing Module, For Clinical Use
Beckman Coulter Inc.
AutoMate 2550 "High Speed" Catalog Number ODL25255; AutoMate 1250 "Standard" Catalog Number ODL25125. Chemistry analyzer sample processing and sorting system.
| Date | Jan 18, 2018 |
| Status | Terminated |
| Reason | A misaligned pipette tip sensor poses the remote risk of cross-contamination when failing to cause a no-tip error, resulting in pipetting patient samples without pipette tips. |
| Root cause | Device Design |
| Action | A recall notification letter was sent to affected US consignees on 1/19/2018 notifying them of the issue. In addition, the recalling firm is instructing field service engineers to inspect the tip present sensor for misalignment at the next service visit. |
| Quantity | 73 units in US |
| Distribution | Worldwide distribution. US nationwide, Algeria, Australia, Austria, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Canada, China, Croatia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Gibraltar, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Korea, Kuwait, Latvia, Lithu |