ClearanceBrief
Recall Z-0403-2018 ยท JAK System, X-Ray, Tomography, Computed

Philips Medical Systems (Cleveland) Inc

Brilliance iCT - Model 728306 Computed Tomography X-ray systems

DateJul 11, 2017
StatusTerminated
ReasonNumerous software issues with Brilliance iCT 4.1.6 version. Issues affect scan start position, data acquisition, exam stop, tube overheat, memory overflow, failure to give error message, length of acquisition, image volume and other functions.
Root causeSoftware design
ActionA Customer Information Letter dated June 29, 2017, was sent to US consignees via certified mail. Philips Key Market representatives will distribute the letter internationally, based on local regulations. Additionally, a Field Service Engineer will contact each site to schedule a time to implement the Field Change Order.
Quantity559 total
DistributionWorldwide distribution, including US nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060937BRILLIANCE VOLUME
Philips Medical Systems (Cleveland), Inc., Jun 5, 2006