Recall Z-0688-2018 ยท DYB Introducer, Catheter

Merit Medical Systems, Inc.

PreludeSNAP Splittable Hydrophilic and non-Hydrophilic Sheath Introducers, Catalog Numbers - PLS-1007, PLS-1008, PLS-1009, PLS-1009.5, PLS-1010, PLS-1010.5, PLS-1011, PLS-2506, PLS-2507, PLS-2508, PLS-2509, PLS-2510, PLS-2510.5, PLSH-1006, PLSH-1009, PLSX-1006, PLSX-1007, PLSX-1009, PLSX-1009.5

DateDec 14, 2017
StatusTerminated
ReasonMerit Medical Systems, Inc. is voluntarily conducting a recall of specific lots of Prelude SNAP Splittable Sheath Introducers due to an intermittent failure of the splittable hub.
Root causeUnder Investigation by firm
ActionCommunication to consignees began on December 14, 2017. Merit requested that consignees stop using the PreludeSNAP devices for human use, discontinue distribution, and return remaining product to Merit Medical.
Quantity60,945 devices
DistributionUS, Australia, EU, Serbia, Israel, Malaysia, South Africa, Bahrain
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K143255Prelude SNAP Splittable Sheath Introducer
Merit Medical Systems, Inc., Apr 9, 2015
K152381Prelude SNAP Splittable Hydrophilic Sheath Introducer
Merit Medical Systems, Inc., Oct 30, 2015