Recall Z-0681-2018 ยท FMI Needle, Hypodermic, Single Lumen

Vygon MFG, Inc., dba/ Churchill Medical Systems, Inc.

19Ga x 1 Huber Needle Set. Product Code: MBS-1910.

DateJul 24, 2017
StatusTerminated
ReasonVygon is conducting a voluntary recall of the products and lots listed in this recall, being initiated because our Tyvek supplier has notified Vygon of issues with their product, specifically the Tyvek may not be coated completely causing gaps in the adhesive used to seal the product packaging. This potential lack of adhesive may affect the sterility of the product.
Root causeUnder Investigation by firm
ActionVygon sent recall letters dated July 24, 2017, to customers via email and fax with read receipts and delivery confirmations. The recall letter informed customers of the reason for recall, product description, code and lot number, and instructions to customers, with a Recall Acknowledgement and Inventory Return Form provided/accompanying the recall letter.
Quantity156
DistributionU.S.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K873957NON-CORING HUBER TYPE NEEDLES
Advantage Medical Systems, Inc., Oct 23, 1987