ClearanceBrief
Recall Z-0737-2018 ยท NJT Immunohistochemistry Reagents And Kits

Ventana Medical Systems Inc

OptiView Amplification Kit (250 Test), Ventana Part Number 860-099, Roche GMMI 06718663001 in vitro diagnostic

DateFeb 27, 2018 (initiated Dec 19, 2017)
StatusTerminated
ReasonIncreased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.
Root causeProcess control
ActionOn Dec 19th all Roche affiliates (the direct consignees) worldwide were notified via Safety Board Notification (SBN) and subsequent Field Safety Notification (EX-US) and UMDC TD00090 (US) SBN and FSN revised Dec 20th due to 3 affected part numbers missed on original letter. The instructions are as follows: "Actions taken by Roche Diagnostics All affected product has been placed on hold. Ventana has reworked all product in its inventory and is in the process of manufacturing new lots for distribution and replacement of customer affected kits. Customers will be notified when corrected produc
Quantity518 units
DistributionWorldwide distribution, including US nationwide, Argentina, Australia, Hong Kong, Indonesia, Philippines, Brazil, Chile, Thailand, Korea, Malaysia, New Zealand, Ecuador, Mexico, Pakistan, Vietnam, Germany, Japan, India, Colombia, Singapore, Uruguay, Peru, Canada, Taiwan, China,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.