ClearanceBrief
Recall Z-0487-2018 ยท DQK Computer, Diagnostic, Programmable

GE Medical Systems Information Technologies, Inc.

GE HEALTHCARE CARESCAPE Central Station (CSCS) software version 2.0.2 The CARESCAPE Central Station is intended to collect information from a network and display this data. This data includes physiological, patient demographic and/or other non-medical information. Physiological parameters and waveforms from monitors and telemetry systems can be displayed, provide alarm annunciation, and can be p

DateJan 25, 2018 (initiated Jan 11, 2018)
StatusTerminated
ReasonCARESCAPE Central Station (CSCS) software version 2.0.2 units may experience unexpected NO COMM (No Communication) and network communication issues after boot-up or system restart.
Root causeSoftware in the Use Environment
ActionThe firm, GE Healthcare, sent an "URGENT MEDICAL DEVICE CORRECTION" letter and email to its consignees on 01/11/2017. The letter described the product, problem and actions to be taken. The consignees were instructed to ensure that all potential users in your facility are made aware of this safety notification and the recommended actions and follow the safety instructions. GE Healthcare will correct all affected products at no cost to you. A GE Healthcare representative will contact you to arrange for the correction. If you have any questions regarding this notification, please contact Techn
Quantity1803 units
DistributionWorldwide Distribution: US (nationwide) and countries of: Algeria, Australia, Austria, Canada, Costa Rica, Czech Republic, Denmark, Egypt, Finland, France, Germany, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Italy, Japan, Jordan, Korea, Republic Of, Kuwait, Lebanon, Mexico, Netherlands, New
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K162012CARESCAPE Central Station V2
GE Healthcare, Sep 15, 2016