Recall Z-0419-2018 ยท DWJ System, Thermal Regulating

Stryker Medical Division of Stryker Corporation

MA3320-PM Premium Adult - SYK Mistral-Air Premium Blankets. MODEL #: MA3302-PM. CATALOG #(s): MA3302-PM

DateJul 21, 2017
StatusTerminated
ReasonIt was identified that the Mistral-Air Premium blankets may be capable of conducting electricity during a defibrillation or cardioversion therapy at higher energy levels (70 Joules or more) when the blanket is applied with the reflective side toward the patient (upside down).
Root causeComponent design/selection
ActionStryker notified customers of the recall by sending Customer Notification and Business Reply Forms, dated July 21, 2017.
Quantity11160
DistributionU.S.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K101705MISTRAL-AIR WARMING SYSTEM
The Surgical Company International B.V., Jul 2, 2010