Recall Z-0413-2018 ยท EKR Plugger, Root Canal, Endodontic

Ormco/Sybronendo

Elements Obturation Unit (EOU) used with respective Buchanan Heat Pluggers Device is an endodontic root canal plugger. It is intended to be used in Dentistry to provide continuous heat at the tip of a dental instrument to test tooth response to thermal stimulus, for tissue cauterization and coagulation, to backfill and downpack gutta percha during Endodontic root canal treatment, and when in Ext

DateDec 27, 2017
StatusTerminated
ReasonBuchanan Heat Pluggers will possibly overheat while used with the Elements Obturation Unit (EOU).
Root causeUnder Investigation by firm
ActionSybron Endo sent an Urgent: Medical Device Recall letter dated December 27, 2017. Customers were informed that Sybron Endo was still evaluating the root cause and would provide additional instructions once the root cause is determined. Distributors were asked to cease distributing any Elements Obturation Units (EOU), confirm if any are in inventory, return any affected devices, provide the notification to customers if further distributed, return any affected devices received from customers, and complete and return the acknowledgement form. Customers were instructed to examine current in
Quantity25,765 total
DistributionUSA (nationwide) distributed to AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. Distributed internationally to Armenia, Australia, Canad
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031664ELEMENTS OBTURATION UNIT
Sybron Endo/Analytic, Apr 28, 2004