Recall Z-0409-2018 ยท OSN Software For Diagnosis/Treatment

Smith & Nephew, Inc.

Smith & Nephew Taylor Spatial Frame Website software, Catalog Number 71070401. A web-based and mobile application software component of a multilateral external fixation system that is intended for fracture fixation, fixation of long bones and for joint fusions and limb lengthening or deformity corrections which involve cutting of the bone.

DateJan 8, 2018 (initiated Dec 13, 2017)
StatusTerminated
ReasonIn rare circumstances, when surgeons use the software to plan the strut adjustment, an inverted image of the frame is generated by the software. The frame will appear inverted in the preview images on the Strut Settings tab.
Root causeUnder Investigation by firm
ActionThe firm initiated their recall by email on 12/13/2017. Users are asked if they notice an inverted frame in the software to repeat the process for the same case and notify a Smith & Nephew representative. In addition, they are asked to double check plannings for strut adjustments and compare with the images printed as described above.
QuantityN/A
DistributionWorldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K142520Smith & Nephew, Inc. Spatialframe.com V5.0 Web-based software
Smith & Nephew, Inc., Nov 20, 2014