Recall Z-0921-2018 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Abbott Laboratories, Inc
ARCHITECT c4000 Cuvette Segment, Part #02P75-01/02P24, a component to the ARCHITECT Clinical Chemistry system. There are 9 segments with each cuvette segment holding 11 cuvettes for a total of 99 cuvettes per reaction carousel. The firm name on the label is Abbott Laboratories, Abbott Park, IL. The Abbott ARCHITECT Clinical Chemistry Systems are designed to perform automated chemistry tests ut
| Date | Jan 18, 2018 (initiated Dec 8, 2017) |
| Status | Terminated |
| Reason | There is a potential to generate falsely depressed patient results in the cuvettes adjacent to the detached segment post for any assay run on the c4000, c8000, and c16000 instrument. |
| Root cause | Device Design |
| Action | The firm, Abbott Laboratories, issued a "Product Correction" letter dated 12/8/2017 via overnight mail on 12/8/2017 to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to adhere to the new instructions listed in Appendix A to avoid damaging the cuvette segments; If a cuvette segment with a detached bottom is identified, replace it before performing any additional testing on your ARCHITECT Clinical Chemistry System; If you have forwarded the product listed above to other laboratories, please inform them of this Product Correction a |
| Quantity | 3,634 ARCHITECT c4000 systems |
| Distribution | Worldwide Distribution: US (nationwide) including Puerto Rico and countries of: Afghanistan, Albania, Algeria, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Burkina Faso, Cameroon, Canada, Cayman |
| FDA record | Open on fda.gov |