Recall Z-0564-2018 ยท DYB Introducer, Catheter

Kalila Medical

Vado Bi-Directional Steerable Sheath 8.8F, REF SS8FMB74, packaged in a lidded tray that is within a pouch which is inserted into a shelf carton, Sterile, Rx. The firm name on the label is Kalila Medical, Campbell, CA.

DateDec 14, 2017
StatusTerminated
ReasonTorn sheath liner material was detected at the proximal end of the device during internal product testing which may be the result of damage incurred during the manufacturing process.
Root causeUnder Investigation by firm
ActionThe recalling firm issued a letter dated 12/14/2017 via FedEx on 12/14/2017.
Quantity261 sheaths
DistributionDistribution was made to DC, FL, GA, MD, MI, NJ, NY, OH, PA, SC, and VA. There was also military/government distribution. There was no foreign distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K162427Vado Bi-Directional Steerable Sheath
Kalila Medical, Inc., Jan 27, 2017