ClearanceBrief
Recall Z-0382-2018 ยท FMI Needle, Hypodermic, Single Lumen

Owen Mumford USA, Inc.

Unifine Pentips 5mm x 31G pen needles, packaged in 30 ct pen needles. Product Usage: Sterile, non-toxic, non-pyrogenic, single use pen needles for the subcutaneous injection of drugs from pen injectors.

DateNov 6, 2017
StatusTerminated
ReasonDevice may potentially be compromised by water in end user packaging. Exposure to liquid may compromise sterility. This may appear as discoloring or yellowing of the device's protective seal.
Root causeEnvironmental control
ActionOwen Mumford sent an Urgent Medical Device Recall Notification letter dated November 6, 2017 via email. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to check inventory for affected product, contact the Customer Service Department to arrange return of affected product, complete and return the response form and to either provide a customer list for Owen and Mumford to notify their customers or confirmation that they will notify their customers if further distributed.
Quantity708 cartons (21,240 pen needles)
DistributionUS Nationwide Distribution in the states of AL, AZ, CA, CO, FL, GA, HI, IA, IL, MA, MI, MN, MO, MS, NC, NH NJ, NY, OH, OK, PA, TN, NX, UT, VA, VT, WA, WV
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K152339Unifine Pentips/ Unifine Pentips Plus
Owen Mumford, Ltd., Jun 24, 2016