ClearanceBrief
Recall Z-0261-2018 ยท OXO Image-Intensified Fluoroscopic X-Ray System, Mobile

Orthoscan, Inc.

Mobile Mini C-arm system Part # 1000-0004-FD. Intended to provide the physician with general fluoroscopic visualization of the patient, including, but not limited to, surgical orthopedic procedures and critical and emergency care procedures.

DateAug 31, 2017
StatusTerminated
ReasonOrthoScan Inc. discovered during investigation of a non-standard work flow a non-conformity of the devices. Specifically, the system software allows a user to activate the Digital Zoom feature in the course of taking a live image exposing a larger x-ray field than can be viewed by the user
Root causeSoftware design
ActionOrthoscan sent an Urgent Recall Notice, dated August 30, 2017, to customers, informing them of the issue and how to address it. Customers are to use the attached list of serial numbers to determine if their device is affected. Customers may continue to use their affected device with caution until the on-site correction has been made. Users should avoid non-conforming workflows by taking the measures outlined in the letter. The firm will contact customers to arrange implementation of the corrective action at no cost to the customer. If there are further questions or concerns, please call
Quantity2012 total
DistributionWorldwide Distribution - USA (nationwide) including Puerto Rico and Hawaii and to the countries of : Australia, Austria, Belgium, Canada, Dominican Republic, Estonia , Finland, France, Germany, Hanover, Hong Kong, Israel, Italy, Japan, Jonesboro, Korea, Malaysia. Netherland, New Zealand, Norway, Pol
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133174ORTHOSCAN FD MINI C-ARM
Orthoscan, Inc., Mar 28, 2014