Recall Z-0351-2018 ยท OIJ Biopsy Needle Guide Kit

Hologic, Inc.

Affirm Lateral Arm Upright Biopsy Accessory. Product Number: ASY-09880

DateJul 7, 2017
StatusTerminated
ReasonThe firm is implementing a product correction involving its Affirm Lateral Arm Upright Biopsy Accessory, ASY-09880. The correction has been initiated due to a component (the Left-hand Guide or the Blue Needle Guide) issue that could impact needle alignment for all the left-approach lateral biopsy procedures. This may have an impact on the position of the needle within the breast, and thus on the aspiration of the intended tissue. This impact is case-dependent and detectable when imagining is used to confirm needle position and/or tissue acquisition during and after a procedure. There are no kn
Root causeComponent change control
ActionLetter
Quantity12 (U.S.)
DistributionUS and OUS
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K161575Affirm Lateral Arm Upright Biopsy Accessory
Hologic, Aug 10, 2016