Recall Z-0492-2018 ยท KHO Fluorometer, For Clinical Use

Tosoh Bioscience Inc

TosohSorter 9tray, Product Code 022931and TosohSorter 19tray, Product Code 022932. Product Usage: The AIA-900 9tray Sorter and AIA-900 19tray Sorter supports the automation of the AIA-900 systems.

DateJun 11, 2017
StatusTerminated
ReasonThis recall has been initiated due to a potential malfunction of the photo interrupter (Pl). The potential malfunction has been attributed to aging and premature deterioration of the Pl collector current.
Root causeNonconforming Material/Component
ActionCustomers were sent a recall notification by email on July 11, 2017 informing them that their local Field Service Engineer would contact them to make arrangements to proactively replace the affected parts. Follow up notifications will be by Federal Express with verification of delivery.
Quantity565 total devices U.S.
DistributionWorldwide Distribution - U.S. Nationwide - Latin America
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.