Recall Z-0438-2018 ยท HSB Rod, Fixation, Intramedullary And Accessories

Zimmer Biomet, Inc.

Affixus Hip Fracture Nail, LH 130 DEG 11MM X 440MM

DateAug 7, 2017
StatusTerminated
ReasonDuring the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Root causeProcess change control
ActionCustomers were notified via FedEx and/or email on approximately 08/07/2017. Instructions include ensure appropriate personnel are notified of the recall, locate and quarantine affected product in inventory, complete and return the Certification of Acknowledgement, return all affected product, and provide the names of hospitals or customers that would have received this product if further distributed.
Quantity638 units
DistributionUS nationwide distribution. Also Portugal and the Netherlands.
FDA recordOpen on fda.gov