Recall Z-0507-2018 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

Microline Surgical, Inc.

The ReNew Reusable Grasper and Dissector Tips are to be used with the ReNew Laparoscopic Handpieces and they are indicated for cutting, grasping, dissecting, and coagulation of tissue in endoscopic and laparoscopic surgical procedures.

DateJul 13, 2017
StatusTerminated
ReasonThe heat-shrink insulation tube may split during autoclave sterilization/re-sterilization. If a split occurs, the heat-shrink could fall off during a surgical procedure and if cautery is needed, there is a potential for burn due to insulation failure.
Root causeDevice Design
ActionThe firm, Microline Surgical, Inc., sent a "URGENT: MEDICAL DEVICE RECALL" letter dated July 13, 2017 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately determine if you are in possession of the product and stop use; quarantine all products both new un-opened Reusable Tips AND Reusable Tips that have been used or are in current use (regardless of whether you have observed the malfunction); If you are a distributor or further distributed this product, we require that you immediately remove affected products from inve
Quantity61,769 units
DistributionWorldwide Distribution-US (Nationwide) and countries of: Australia, Brazil, Canada, Chile, China, Colombia, El Salvador, Hong Kong, India, Indonesia, Malaysia, Mexico, Panama, Peru, Argentina, South Korea, Sri Lanka, Taiwan, Thailand, Uruguay & Vietnam.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K962119RE-NEW LAPAROSCOPIC INSTRUMENTS
Microline Pentax, Inc., Aug 9, 1996
K974066RE-NEW FORCEPS LAPAROSCOPIC SURGICAL TIPS
Microline Pentax, Inc., Jan 22, 1998