Recall Z-0387-2018 ยท DHB System, Test, Radioallergosorbent (Rast) Immunological

Phadia US Inc

Phadia 1000 Instrument, Article Number 12-3800-01.

DateJul 5, 2017
StatusTerminated
ReasonThe "Retry" command does not function properly which could cause a shortage of Wash and Rinse solution and affect assay performance and test results.
Root causeDevice Design
ActionThe firm sent letters with a return response form to customers. The firm instructed customers to immediately choose the "stop" command if presented with the 7-101 Liquid Sensor Error. Results from the assay should not be approved. A revision of the current software will have a mandatory installation soon.
Quantity70 units
DistributionNationwide
FDA recordOpen on fda.gov