ClearanceBrief
Recall Z-0329-2018 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Encore Medical, Lp

EMPOWR Porous Knee System TIBIAL PUNCH, LARGE, REF 801-05-204, NON STERILE, Qty 1 For use in orthopedic surgery.

DateDec 12, 2017
StatusTerminated
ReasonThe Locking Punch Guide was not sufficiently guiding the large press fit punch, causing the implant to be placed 2-3mm more posterior than intended.
Root causeDevice Design
ActionThe field safety notice was disseminated by email on 12/12/2017. The notice stated the following: "DJO Sales Agents Action: Please review the bulletin that is provided with this notice. This bulletin contains new instructions on the use of the Locking Punch Guide and Punch. Please ensure your surgeons are familiar with this information. DJO Surgical then requires that you complete the attached acknowledgment and email it to teffany.hutto@djoglobal.com. Please have this form returned by December 15, 2017 Once redesigned guides are available, they will be exchanged through a separate fi
Quantity2 units
DistributionOne medical device distributor in California.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K100900MOTIVATION KNEE SYSTEM
Encore Medical L.P., Jun 18, 2010