Recall Z-0320-2018 ยท HAW Neurological Stereotaxic Instrument

Brainlab AG

Brainlab Navigation System Spine & Trauma 3D with component Brainlab DrapeLink Registration Kit for C-arm Zeihm Vision RFD: Adapter Left

DateNov 16, 2017
StatusTerminated
ReasonDue to a human error at a supplier, an unknown number of devices within one specific production lot of Brainlab DrapeLink Adapters (Left/Right) were not manufactured according to defined specifications.
Root causeNonconforming Material/Component
ActionCustomers were notified, via email on approximately 11/16/2017. Instructions included to identify all affected devices on hand, cease use of the affected device, and a Brainlab Customer Support Representative will contact the customer to arrange a service visit to replace the item
Quantity80 devices total
DistributionDistributed in the US to Kansas. Distributed internationally to Australia, Finland, France, Germany, Hong Kong, Ireland, Japan, Netherlands, Norway, Sweden, Switzerland, Taiwan, Turkey, United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K070106VV FLUORO 3D
Brainlab AG, Apr 23, 2007