Recall Z-0348-2018 ยท MKJ Automated External Defibrillators (Non-Wearable)
Philips Electronics North America Corporation
ECG Out Cable - Heartstart MRx M1783A/M5526A 12-pin Sync cables All units of Philips sync cables model M1783A and M5526A as well as Philips ECG out cables model 989803195641. Manufactured - September 4, 2002 October 8, 2014
| Date | Jul 17, 2017 |
| Status | Terminated |
| Reason | Issue with ECG out cables. When a Philips monitor/defibrillator is receiving an ECG signal from an auxiliary bedside monitor via a sync cable or ECG out cable, the following can occur if the monitor/defibrillator experiences interference from electrical fast transients (EFTs): " On the HeartStart MRx and HeartStart XL, EFT noise may be misinterpreted as an R-wave. " On the HeartStart XL+, EFT noise can disable ECG monitoring and potentially interrupt demand mode pacing*. *Note: It is contrary to the XL+ Instructions for Use to perform demand mode pacing while using the ECG out cable or obt |
| Root cause | Device Design |
| Action | Philips is asking customers to follow the Action to be Taken by Customer/User section of the Urgent Medical Device Correction notification/Field Safety Notice: Customers should remove affected sync cables and ECG out cables from service and destroy them, documenting such action on the attached reply form, which should be returned to Philips. Customers should retain a copy of the reply form. The customer should follow the Instructions for Use that describe the recommended approach for performing cardioversion with the XL+ and MRx monitor/defibrillators. The ECG trunk cable can be disconnect |
| Quantity | 15811 |
| Distribution | 1067 US customers 163 Canada, 917 International |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K031187 | HEARTSTART MRX, MODEL M3535A Philips Medical Systems, Aug 27, 2003 |
| K040404 | HEARTSTART MRX, MODEL M3535A Philips Medical Systems, Inc., May 4, 2004 |
| K051134 | HEARTSTART MRX WITH Q-CPR OPTION, MODELS M3535A OR M3536A Philips Medical Systems, Sep 8, 2005 |
| K061707 | HEARTSTART MRX, INVASIVE PRESSURE & TEMPERATURE OPTIONS Philips Medical Systems, Aug 11, 2006 |
| K062233 | HEARTSTART MRX MONITOR/DEFIBRILLATOR Philips Medical Systems North America, Inc., Nov 22, 2006 |