Recall Z-0278-2018 ยท MSX System, Network And Communication, Physiological Monitors

Philips Electronics North America Corporation

Automated External Defibrillator - HeartStart MRx Monitor/Defibrillator

DateSep 28, 2017
StatusTerminated
ReasonPhilips MRx devices have an EtCO2 module that may contain contamination; this contamination (will/may) cause the device to be unable to display the EtCO2 or CO2 waveform (capnogram).
Root causeNonconforming Material/Component
ActionThe Philips Urgent Medical Device Correction notification/ Field Safety Notice informs customers of the issue, identifies details of the units affected, gives instructions on actions to be taken by the customer and identifies what action Philips plans to take to remedy the issue. The correction will consist of providing a hardware correction, free of charge, for all units affected by this issue.
Quantity20 units ( 18 domestic, 2 foreign)
Distribution20 units affected. 18 domestic, 2 foreign.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031187HEARTSTART MRX, MODEL M3535A
Philips Medical Systems, Aug 27, 2003
K040404HEARTSTART MRX, MODEL M3535A
Philips Medical Systems, Inc., May 4, 2004
K051134HEARTSTART MRX WITH Q-CPR OPTION, MODELS M3535A OR M3536A
Philips Medical Systems, Sep 8, 2005
K061707HEARTSTART MRX, INVASIVE PRESSURE & TEMPERATURE OPTIONS
Philips Medical Systems, Aug 11, 2006
K062233HEARTSTART MRX MONITOR/DEFIBRILLATOR
Philips Medical Systems North America, Inc., Nov 22, 2006
K063375HEARTSTART MRX, MODELS M3535A OR M3536A
Philips Medical Systems, Jan 11, 2007