ClearanceBrief
Recall Z-0287-2018 ยท IYE Accelerator, Linear, Medical

Accuray Incorporated

TomoHD(R) System, Model Number 1018283 Product Usage: The TomoTherapy treatment system is intended to be used as an integrated system for t he planning and precise delivery of radiation therapy, stereotactic radiotherapy, or stereotactic radiosurgery to tumors or other targeted tissues while minimizing the delivery of radiation to vital healthy tissues. The megavoltage x-ray radiation is delivere

DateNov 17, 2017
StatusTerminated
ReasonUncontrolled couch Z-axis movement (descent)
Root causeSoftware design
ActionAccuray sent an Urgent Field Safety Notice letter dated November 8, 2017, to affected customers. The letter identified the affected product, problem and actions to be taken. The letter stated the following: "Accuray(R) will correct all affected High-Performance Patient Couches. An Accuray representative will contact you to arrange a time to update your system. The update includes the installation of TomoTherapy(R)' H(TM) Series software version 2.1.2 (Hi Art(R) vS.1.2) on your system, as well as the replacement of cables within the couch subsystem." For questions contact Accuray Customer Su
Quantity118 units
DistributionWorldwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K121934TOMOTHERAPY TREATMENT SYSTEM
Accuray Incorporated, Aug 29, 2012