Recall Z-0286-2018 ยท IYE Accelerator, Linear, Medical
Accuray Incorporated
Hi-Art(R) System, Model Number H-000-0003 Product Usage: The TomoTherapy treatment system is intended to be used as an integrated system for t he planning and precise delivery of radiation therapy, stereotactic radiotherapy, or stereotactic radiosurgery to tumors or other targeted tissues while minimizing the delivery of radiation to vital healthy tissues. The megavoltage x-ray radiation is deliv
| Date | Nov 17, 2017 |
| Status | Terminated |
| Reason | Uncontrolled couch Z-axis movement (descent) |
| Root cause | Software design |
| Action | Accuray sent an Urgent Field Safety Notice letter dated November 8, 2017, to affected customers. The letter identified the affected product, problem and actions to be taken. The letter stated the following: "Accuray(R) will correct all affected High-Performance Patient Couches. An Accuray representative will contact you to arrange a time to update your system. The update includes the installation of TomoTherapy(R)' H(TM) Series software version 2.1.2 (Hi Art(R) vS.1.2) on your system, as well as the replacement of cables within the couch subsystem." For questions contact Accuray Customer Su |
| Quantity | 250 units |
| Distribution | Worldwide Distribution |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K121934 | TOMOTHERAPY TREATMENT SYSTEM Accuray Incorporated, Aug 29, 2012 |