Recall Z-0324-2018 ยท FGE Stents, Drains And Dilators For The Biliary Ducts

Cordis Corporation

Cordis S.M.A.R.T. Flex Biliary Stent System Product Usage: Is indicated for use in the palliation of malignant strictures in the biliary tree.

DateNov 15, 2017
StatusTerminated
ReasonPotential for cracked luer hubs
Root causeNonconforming Material/Component
ActionOn November 15, 2017 the firm sent letter to their customers indicating the following: The purpose of this communication is to inform you of a potential leakage scenario with the S.M.A.R.T. Flex Biliary Stent System. The leak may occur due to potentially cracked hubs in some units from within the 29 lots listed in this letter. Inspection details: As stated in the Instructions for use, users are warned that the device should not be used if the outer sheath cannot be flushed. Inspect for leakage during preparation of the device. Leakage may be more likely in the affected product due to cracks i
Quantity511
DistributionWorldwide Distribution - US Nationwide in the states of MI, AL, AR, CA, CO, FL, GA, IL, IA, LA, MS, NJ, NC, OH, OK, OR, PA, TN, TX, VA, WV, WI, MI, NY, MA, NV, RI, IN, WA, AZ, DC, Puerto Rico and the country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K121125FLEXSTENT BILIARY SELF-EXPANDING STENT SYSTEM
Flexible Stenting Solutions, Inc., Jul 6, 2012