Recall Z-0515-2018 ยท CHP Radioimmunoassay, Estradiol

Siemens Healthcare Diagnostics, Inc

ADVIA Centaur Calibrator 30, 2 pack, 6 pack, and Ref 2 pack. Siemens Material Number (SMN): 10379810, 10379811, 10697756.

DateJun 7, 2017
StatusTerminated
ReasonSiemens Healthcare Diagnostics has confirmed a negative bias for ADVIA Centaur Enhanced Estradiol (eE2) on the ADVIA Centaur, ADVIA Centaur XP and ADVIA Centaur XPT systems when calibrating with Calibrator 30 kit lots ending in 21 (C3021) as compared to Calibrator 30 kit lots ending in 20 (C3020).
Root causeUnder Investigation by firm
ActionAn Urgent Medical Device Recall letter dated June 7, 2017 was distributed to all customers in the United States and an Urgent Field Safety Notice was distributed to all customers outside the United States who have received ADVIA Centaur Systems Calibrator 30 kit lot ending in 21(C3021). These letters instruct customers who operate the ADVIA Centaur, ADVIA Centaur XP and ADVIA Centaur XPT systems to discontinue use of C3021 . The letters inform customers that Calibrator 30 kit lots ending in 20 (C3020) and 23 (C3023) are not impacted and can be used. The letters inform customers who operate ADV
Quantity5992 units (US and OUS)
DistributionUS and OUS.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K100293ADVIA CENTAUR CALIBRATOR 30, MODELS 10379811/10379810
Siemens Heatlhcare Diagnostics, Inc., Mar 19, 2010