ClearanceBrief
Recall Z-0246-2018 ยท LJS Catheter, Intravascular, Therapeutic, Long-Term Greater Than 30 Days

Angiodynamics Inc. (Navilyst Medical Inc.)

Bio-Stable 5F DL-55CM MST-70 KIT Non-Valved PG, UPN H965458710, Catalog No. 45-871 The Xcela with PASV, BioFlo PICC with ENDEXO Technology, and BioFlo PICC with ENDEXO and PASV Valve Technology are indicated for short or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications, nutrients; the sampling

DateNov 28, 2016
StatusTerminated
ReasonNavilyst Medical, Inc. (NMI) the manufacturer of the Xcela PICC with PASV, BioFlo PICC with PASV and the BioFlo PICC, conducted this recall to the end user level based on information received from Greatbatch Medical, the manufacturer of the ViaPeel PTFE Peelable Introducer. Greatbatch Medical determined that the products listed in their 11/11/2016 Recall Notification have the potential for the handles to detach from the sheath during use. NMI has confirmed that the affected sheaths, Greatbatch Model Number 10890-006, had been included in packaged Xcela and BioFlo PICC Kits.
Root causeComponent design/selection
ActionThe firm, Navilyst Medical, sent an "URGENT VOLUNTARY MEDICAL DEVICE RECALL IMMEDIATE ACTION REQUIRED" letter dated 11/28/2016 by issuing recall notifications to consignees via Federal Express. The letter described the product, problems and actions to be taken. The Consignees were instructed to 1. IMMEDIATELY o Stop using the product subject to recall. o Remove any affected (recalled) product from your inventory (whether in labs, Central Supply, Shipping and Receiving or ANY other location). o Segregate this product in a secure location for return to Navilyst Medical, Inc. (an AngioDynamics C
Quantity4591 units in total
DistributionWorldwide Distribution: US (Nationwide) and countries of:; Canada, Great Britain, Netherlands, South Australia and Spain.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K150448NMI PICC III and NMI PICC IV
Navilyst Medical, Inc., Mar 19, 2015