ClearanceBrief
Recall Z-0309-2018 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Beckman Coulter Inc.

Access 2 Immunoassay System, Remanufactured, Catalog Number 386220. In vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids.

DateNov 6, 2017
StatusTerminated
ReasonPotential for minor skin burn if the tool contacts the reagent storage interface board (B79917) when performing a pipettor alignment.
Root causeDevice Design
ActionAn Urgent Medical Device Recall letter, dated November 6, 2017, were sent to the affected customers via mail. The letter identifies the affected device, the reason for recall, impact, and actions to be taken. The firm states that customers may continue to operate their system normally. Customers are instructed to contact Technical Support if a device on the Access 2 instrument needs alignment. The alignment tool is only to be used by the Beckman Coulter service personnel and Beckman Coulter trained biomedical engineers. A Beckman Coulter service rep will contact customers to schedule a s
Quantity3 units
DistributionWorldwide Distribution -- US, Argentina, Canada, Chile, China, Columbia, Costa Rica, France, Georgia, Germany, Ghana, Greece, Hong Kong, India, Italy, Japan, Kazakhstan, Mexico, Morocco, Myanmar, Nepal, Philippines, Poland, Russia, Rwanda, Saudi Arabia, Slovakia, South Africa, South Korea, Thailand
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K922823ACCESS THEOPHYLLINE IMMUNOASSAY REAGENTS
Bio-Rad Laboratories, Inc., Aug 5, 1992