ClearanceBrief
Recall Z-0238-2018 ยท GKZ Counter, Differential Cell

Beckman Coulter Inc.

UniCel DxH Slidemaker Stainer, Part Number 75222, for use with automated hematology analyzer

DateNov 3, 2017
StatusTerminated
ReasonThe shield for the Single-Tube Station on the DxH instrument may interfere with sample aspiration in the open vial position.
Root causeDevice Design
ActionThe firm notified its consignees of the recall on 11/03/2017 by letter. The letter stated the issue and provided the following instructions: "To avoid a biohazardous spill, do not use the right (green) position of the Single-Tube Station for sample aspiration until the shield is removed. You may remove the shield following the instructions in Attachment 1 or contact your local Beckman Coulter Representative for further assistance."
Quantity3 units
DistributionCA, AR, HI, CA, OH, WA, MD, Canada, Japan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K140911UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM/ UNICEL DXH SLIDEMAKER STAINER COULTER CELLULAR ANALYSIS SYSTEM
Beckman Coulter, Inc., Sep 5, 2014