ClearanceBrief
Recall Z-0896-2018 ยท LKK Pump, Infusion, Implanted, Programmable

Medtronic Neuromodulation

Medtronic Synchromed II Model 8637 supplied in 20mL and 40 mL reservoir size and a catheter. Product Usage: The implantable Model 8637 SynchroMed II programmable pump is part of an infusion system that stores and delivers a prescribed drug to a specific site. The implanted infusion system consists of a Model 8637 SynchroMed II pump, supplied in 20 ml and 40 ml reservoir sizes, and a catheter. In

DateAug 9, 2017
StatusOpen, Classified
ReasonMedtronic implemented a final design change intended to address the motor stall due to corrosion and is recalling devices with the previous design that are not implanted.
Root causeDevice Design
ActionMedtronic sent an Urgent Medical Device Recall letter dated September 2017, to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete and return the enclosed Customers Confirmation form. For questions contact Technical Services at 1-888-854-0978.
Quantity27,547 US and 13,025 OUS
DistributionUS Nationwide Distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OR, OK, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.