Recall Z-0401-2018 ยท DWA Control, Pump Speed, Cardiopulmonary Bypass

Sorin Group Deutschland GmbH

CP5 Centrifugal Pump, made up of the drive unit, the emergency system and the pump control panel. It used with an Electrical Remote-Controlled (ERC) tubing clamp and a bubble sensor.

DateOct 30, 2017
StatusTerminated
ReasonPossibility that the automatic line closure by the Electronic Remote-Controlled (ERC) clamp after detection of an air bubble could take longer than is intended by design in the CP5 system.
Root causeDevice Design
ActionConsignees were notified of the recall via USPS trackable mail, beginning on 10/30/2017. The letter informed customers Instructions included to continue following the Instructions for use, including use of an arterial filter on the arterial line, and that a LivaNova representative will contact them to schedule an upgrade to their device.
Quantity1,318 units
DistributionDistributed domestically to Distributed internationally to Algeria, Australia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Canada, China, Colombia, Croatia, Czech Republic, Ecuador, Estonia, Finland, France, Germany, Hong Kong, Iraq, Italy, Japan, Jordan, Kuwait, Libya, Lithuania, Luxembourg,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K112225SORIN CENTRIFUGAL PUMP 5 (CP5)
Sorin Group Deutschland GmbH, Sep 20, 2011