Recall Z-0323-2018 ยท MEA Pump, Infusion, Pca

CareFusion 303, Inc.

Alaris PCA Pump, Model 8120. The firm name on the label is CareFusion. The Alaris Syringe Pump module and Alaris PCA module are infusion pump modules for the Alaris System. The syringe pump and PCA pump delivers medication or fluids in a manner similar to current syringe pumps on the market. The modules are indicated for use on adults, pediatrics and neonates for continuous or intermittent deliv

DateNov 17, 2017
StatusTerminated
ReasonThe syringe plunger grippers may fail to automatically close around the syringe plunger press when the gripper control knob has been closed.
Root causeNonconforming Material/Component
ActionThe recalling firm issued "Medical Device Safety Notification, dated November 17, 2017, via UPS. The letter describes the potential issue, affected products, potential risks, and required action for the users. The actions noted that clinicians can continue to use the device and follow the Alaris System User Manual for proper loading of the syringe. Per the manual, clinicians should ensure the plunger grippers lock and hold the syringe plunger in place. The Biomedical Engineering can detect if the device has this issue and if the issue described occurs, the user is to contact BD Support
Quantity73,611
DistributionWorldwide Distribution - USA (nationwide) Distribution. There was also government and military distribution. Foreign distribution was made to Canada, United Arab Emirates, Bahrain, Canada, Switzerland, United Kingdom, Kuwait, Mexico, Malaysia, The Netherlands, New Zealand, Qatar, Saudi Arabia, Taiwa
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K032233MEDLEY PCA MODULE, MODEL 8120
Alaris Medical Systems, Inc., Sep 9, 2003