Recall Z-0189-2018 ยท OBJ Catheter, Ultrasound, Intravascular

Volcano Corporation

Volcano Imaging System s5i REF 807400001; CORE Mobile Imaging System (120V) REF 400-0100.01; CORE Mobile Imaging System Refurbished REF 400-0100.01-R; CORE Mobile Imaging System (240V) REF 400-0100.07; CORE Mobile Imaging System Refurbished REF 400-0100.07-R; CORE Mobile Imaging System (100V) REF 400-0100.08; CORE Mobile Imaging System Refurbished REF 400-0100.08-R; CORE Imaging System REF 400-010

DateNov 3, 2017
StatusTerminated
ReasonOn certain Impacted Systems, an unexpected Microsoft Windows Security dialog may appear during use, and the users response to the dialog may adversely affect the subsequent operation of the device.
Root causeSoftware design
ActionBeginning the week of November 7, 2017, all affected US customers will be sent a field correction notification letter along with a Technical Service Bulletin (TSB) to explain how to assure the misconfiguration does not affect their system performance. For Distribution outside the US, the same letter and TSB will be sent to the initial consignee and they will execute the delivery of the Field Correction Notice and will inform the Recalling firm of the responses.
Quantity1166 Units
DistributionWorldwide distribution including US nationwide and Canada, AE, AT, AU, BD, BE, BG, BR, CH, CL, CN, CY, DE, DK, EG, ES, FR, GB, HU, IE, IL, IN, IT, JO, JP, KW, KR, LB, MA, MX, MY, NL, NO, PA, PK, PL, PT, QA, RO, RU, SA, SE, SG, SK, Sri Lanka, TH, TW, UAE, UK, ZA,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K071554VOLCANO S5/S5I INTRAVASCULAR IMAGING AND PRESSURE SYSTEM
Volcano Corporation, Jan 4, 2008
K123898VOLCANO S5/S5I INTRAVASCULAR ULTRASOUND IMAGING AND PRESSURE SYSTEM
Volcano Corporation, Jan 14, 2013