ClearanceBrief
Recall Z-0580-2018 ยท BSZ Gas-Machine, Anesthesia

GE Medical Systems China Co., Ltd.

Carestation 620 A2, Carestation 650 A2, and Carestation 650c A2

DateOct 11, 2017
StatusOpen, Classified
ReasonPossible incomplete seal between the CO2 absorber and the breathing circuit CO2 bypass port assembly of the Carestation 600 Series systems.
Root causeOther
ActionOn approximately 10/11/2017 customers were notified of the recall via letter. Instructions include that if elevated FiCO2 is observed, to increase the flow of fresh gas to reduce the volume of patient gas that is rebreathed. If the FiCO2 cannot be adequately reduced with this action, consider switching to another anesthesia delivery device. GE Healthcare is planning to release revised parts that minimize the likelihood of incomplete gas flow through the CO2 absorbent canister. GE Healthcare will correct all affected devices by installing updated parts at no cost to the customer. The upd
Quantity5,148 devices
DistributionDistributed domestically to AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NV, NY, OH, OK, PA, TN, TX, UT, VA, VT, WA, WI, WV. Distributed internationally to Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Boli
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.