ClearanceBrief
Recall Z-2050-2018 ยท LLZ System, Image Processing, Radiological

NordicNeuroLab AS

nordicBrainEx 2.0 provides analysis and visualization capabilities of dynamic MRI data of the brain, presenting the derived properties and parameters in a clinically useful context.

DateSep 17, 2014
StatusTerminated
ReasonAn error was discovered in the interpretation of certain DICOM header tags that may lead to incorrect orientation labeling, and thus and indirect left-right, up-down or anterior-posterior flipping of images.
Root causeSoftware design
ActionThe firm notified their consignees by email on 09/17/2014.
Quantity22 licenses
DistributionUSA, Sweden, Spain, Russia, Hungary, India, Norway, Iceland, Italy, South Korea, United Kingdom
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133910NORDICBRAINEX
Nordicneurolab, Apr 4, 2014