Recall Z-2044-2018 ยท LLZ System, Image Processing, Radiological

NordicNeuroLab AS

nordicICE 2.3.14 Image processing software package used by trained professionals, including physicians and medical technicians.

DateSep 12, 2014
StatusTerminated
ReasonAn error was discovered in the interpretation of certain DICOM header tags that may lead to incorrect orientation labeling, and thus and indirect left-right, up-down or anterior-posterior flipping of images.
Root causeSoftware design
ActionThe firm notified their consignees by e-mail on 09/12/2014. Customers were informed of the reason for the recall, when the problem occurs, and what actions to take. Customers can perform a simple test by checking to see if there is a mismatch. Also, it is recommended that customers using BOLD and DTI, or other functionalities or workflows involving MPR switch to nordicBrainEx. Questions should be directed to NordicNeuroLab Customer Service at customerservice@nordicneurolab.com or 1-262-337-2909.
Quantity97 licenses
DistributionWorldwide Distribution -- USA, Australia, Denmark, China, France, Japan, Malaysia, Netherlands, USA, Poland, Sweden, Spain, Russia, India, Norway, Singapore, South Korea, United Kingdom, and Slovenia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082441NORDICICE SOFTWARE
Nordicimaginglab AS, Nov 21, 2008