Recall Z-0195-2018 ยท FRN Pump, Infusion
ICU Medical Inc
ICU Mednet(TM) Medication Management Suite 6.1 and 6.21, Product Code/List Numbers: (a) 16037-64-01; MedNet 6.1 b) 16037-64-02; MedNet 6.1 c) 16037-64-03; MedNet 6.1 d) 16037-64-04; MedNet 6.21 The ICU Medical MedNet Medication Management Suite (MMS) is intended to facilitate networked communication between MMS compatible computer systems and MedNet compatible infusion pumps. The MMS provides t
| Date | Oct 30, 2017 |
| Status | Terminated |
| Reason | Issue 1: The MedNet Meds 6.1 and 6.21 programs, under certain conditions, can change the piggyback medication entry set settings for existing defined medication entries. This includes (a) whether a medication may be piggybacked and (b) whether an infusion may be interrupted by a piggyback. This only occurs when: 1. Medication entries are built with the same displayed name and diluent amount AND; 2. The piggyback or interruptible settings differ from one CCA to another AND; 3. The drug library is exported then imported into the MedNet Meds program. The changed drug library may then be finalize |
| Root cause | Under Investigation by firm |
| Action | ICU Medical sent an Urgent : Medical Device Correction letter dated October 30, 2017, Please Inform potential users in your organization of this notification and complete the attached response form and return it to the fax number or e-mail address on the form,even if you do not have the affected product. If you have distributed the product further, notify your accounts that received the product identified above of this notification and ask them to contact Stericycle at 1-855-215-5460 (M-F,8am-5pm ET) to receive a reply form. ICU Medical is committed to patient safety and is focused on prov |
| Quantity | 108 installations |
| Distribution | Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Canada, Australia and Philippines |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K141102 | HOSPIRA MEDNET MEDICATION MANAGEMENT SUITE Hospira, Inc., Jul 8, 2014 |