Recall Z-0489-2018 ยท MKJ Automated External Defibrillators (Non-Wearable)

Philips North America, LLC

MRx Defibrillator Model # M3535A & M3536A, UDI 00884838000018(21)US00601969 Product Usage: The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used

DateFeb 8, 2018 (initiated Jul 12, 2017)
StatusTerminated
ReasonAffected units may have a defective component that is intended to protect the internal circuitry from high voltages during shock delivery. The defective component may also prevent delivery of shocks in AED mode.
Root causeNonconforming Material/Component
ActionPhilips Healthcare sent an Urgent Medical Device Correction Notification/Field Safety Notice letter dated June, 2017 to their customers. The letter identified the affected product, problem, hazard involved and actions to be taken. Philips is requesting that customers report if their device fails to function as intended. Philips will contact all customers for device replacement once parts are available. If you need further information or support concerning this action, please contact your local Philips representative.
Quantity71 units
DistributionNo U.S. distribution, Internationally to: India, Ireland, Japan, Republic of Korea, Malaysia, Mexico, South Africa & United Kingdom
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031187HEARTSTART MRX, MODEL M3535A
Philips Medical Systems, Aug 27, 2003
K040404HEARTSTART MRX, MODEL M3535A
Philips Medical Systems, Inc., May 4, 2004
K051134HEARTSTART MRX WITH Q-CPR OPTION, MODELS M3535A OR M3536A
Philips Medical Systems, Sep 8, 2005
K061707HEARTSTART MRX, INVASIVE PRESSURE & TEMPERATURE OPTIONS
Philips Medical Systems, Aug 11, 2006
K062233HEARTSTART MRX MONITOR/DEFIBRILLATOR
Philips Medical Systems North America, Inc., Nov 22, 2006