Recall Z-0516-2018 ยท LXH Orthopedic Manual Surgical Instrument

Zimmer Biomet, Inc.

DRILL BIT MINI QUICK CONNECT 1.0MM, REF 2312-20-200 & 2312-20-202, NON STERILE

DateJun 12, 2017
StatusTerminated
ReasonA raw material anomaly was discovered in four lots of raw material and has the potential to be on or below the surface of the device components which can increase the risk of instrument fracture
Root causeNonconforming Material/Component
ActionOn June 12, 2017, Zimmer Biomet distributed Urgent Medical Device Recall Removal and Certificates of Acknowledgement notices to their customers via email and FedEx. Risk Manager Customer Responsibilities are the following: 1. Review the notification and ensure affected personnel are aware of the contents. 2. Assist your Zimmer Biomet sales representative and quarantine all affected product. 3. Your Zimmer Biomet sales representative will remove the affected product from your facility. 4. Complete and return the Certificate of Acknowledgement via email to: corporatequality.postmarket@zimmerbiom
Quantity1,524 units
DistributionAK, AL, AR, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA MD, MI, MN, MO, MT, NC, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA & WV. Internationally to: Australia, Chile, Cost Rica, Ecuador, Netherlands, Panama & Switzerland
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.