Recall Z-0072-2018 ยท BSZ Gas-Machine, Anesthesia

Spacelabs Healthcare, Ltd.

Arkon Anesthesia Delivery System, Model 99999, and software version 2.61 and upgrade kit 050-9043-00 Rev G. The Spacelabs Arkon Anesthesia Workstation is intended for use in the hospital environment and operating room. It may be used for the delivery of oxygen, air and nitrous oxide in a controlled manner to various patient breathing circuits with or without the use of mechanical ventilator, and

DateNov 21, 2017 (initiated Oct 11, 2017)
StatusTerminated
ReasonArkon Anesthesia Workstation, with software version 2.61, experienced failure in mechanical ventilation, oxygen and anesthetic gas delivery, with concurrent failure of the display unit that resulted in a blank screen without audible or visible alarms.
Root causeSoftware design
ActionSpacelabs Healthcare sent an Urgent Medical Device Correction letter dated October 24, 2017. The letter identified the affected product problem and actions to be taken. For questions call 1-800-522-7025 and select 4 for Technical Support.
Quantity110 units
DistributionNationwide Distribution to the states of NM, WY, FL, NC, MS, CO, CT, ME, and AL.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K113051SPACELABS ARKON ANESTHESIA WORKSTATION