Recall Z-0084-2018 ยท JKA Tubes, Vials, Systems, Serum Separators, Blood Collection
Becton Dickinson & Company
BD Vacutainer Sodium Polyanetholesulfonate 5.95 mg Sodium Chloride 14.4 mg Blood Collection Tubes for Microbiological Studies (8.3 mL * 16x100mm) Catalog No. 364960
| Date | Jan 27, 2017 |
| Status | Terminated |
| Reason | After a receiving a customer complaint for incorrect labeling, BD has confirmed that a portion of SPS tubes associated with catalog 364960, lot 6090812 were incorrectly labeled at the case and shelf pack level. |
| Root cause | Error in labeling |
| Action | BD notified their customers on 1/27/2017 via email and/or UPS 2nd Day Mail. Stericycle is handling returns. |
| Quantity | 90,000 tubes |
| Distribution | Nationwide, Canada |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.