Recall Z-0114-2018 ยท FLL Continuous Measurement Thermometer

3M Company - Health Care Business

3M Bair Hugger(TM) Normothermia System, Temperature Monitoring System Sensors Model 360 (Part Number 36000)

DateOct 17, 2017
StatusTerminated
ReasonDuring a recent investigation, 3M confirmed that a programming translation error could occur in a small amount of the sensors that could lead to a temperature readout that is lower than the patients actual temperature.
Root causeSoftware design
ActionThe firm initiated their recall on 10/17/2017 via mail. International subsidiaries were notified via email, also on October 17, 2017.Distributors were asked to notify their customers and all direct consignees were directed to dispose of the product on site.
Quantity153 cases (3825 sensors)
DistributionCA, MI, IL, MO, MN, GA, and Canada, Switzerland
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K1204123M SPOTON TEMPERATURE MONITORING SYSTEM
Arizant Healthcare, Inc., May 30, 2012