Recall Z-0194-2018 ยท GEX Powered Laser Surgical Instrument

Monteris Medical Corp

MONTERIS MEDICAL NEUROBLATE SYSTEM, used for MRI-guided neurosurgical ablation.

DateMar 20, 2018 (initiated Oct 5, 2017)
StatusTerminated
ReasonThe firm received complaints that involved an unanticipated interaction between a NeuroBlate system and 1.5 Tesla (1.ST) MRI systems.
Root causeDevice Design
ActionThe firm issued advisory notice 10/05/2017. The notice discussed complaints received that involved an unanticipated interaction between a NeuroBlate system and 1.5 Tesla (1.5T) MRI systems. The notice provided added recommendations for all NeuroBlate users to mitigate the anomaly.
Quantity52 systems
DistributionNationwide and Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K171255Monteris Medical NeuroBlate System
Monteris Medical, Inc., Jul 25, 2017