Recall Z-0074-2018 ยท LQL Gram Positive Identification Panel

Biomerieux Inc

VITEK2 GP-ID, REF 21342 IVD

DateOct 11, 2017
StatusTerminated
ReasonCustomers have reported, and an internal investigation has confirmed, occurrences of atypical negative O129R reactions for Enterococcus casseliflavus ATCC)R) 700327(TM) in association with three (3) VITEK(R) 2 GP ID Lots, causing failure of the Quality Control (QC) testing for the VITEK(R) 2 GP ID card. In accordance with the VITEK(R) 2 GP ID Instructions For Use, the expected O129R reaction for ATCC(R) 700327(TM) strain is positive. As a precaution, bioMrieux is also including in the scope of this notice all GP ID card lots manufactured with the same raw materials.
Root causeUnder Investigation by firm
ActionConsignees (Subsidiaries) of bioMrieux, impacted by the identified issue, were notified of the FSCA and corrective action (FSCA notification) electronically on 10/11/2017. The direct consignees (subsidiaries and distributors) have until 11/11/2017 to notify their respective impacted users. The directions to the user are as follows: "Please implement the following actions at this time: We request you take the following action at this time: " Please confirm this letter has been distributed and reviewed by all appropriate personnel within your organization. " Please refrain from using the refere
Quantity5672 cartons
Distributionworldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K952095VITEK GRAM POSITIVE IDENTIFICATION CARD (GPI)
Biomerieux Vitek, Inc., Mar 12, 1996