Recall Z-0166-2018 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Abbott Laboratories, Inc
Tubing, Peristaltic Head Part Number 7-35009685-01 for use in clinical chemistry systems.
| Date | Sep 29, 2017 |
| Status | Completed |
| Reason | The Peristaltic Head tubing placed on the ARCHITECT c4000, c8000, and c16000 has the possibility to leak due to a manufacturing issue associated with the tubing connector. |
| Root cause | Component design/selection |
| Action | All worldwide ARCHITECT Clinical Chemistry System customers with impacted instruments will receive the Product Correction Letter dated, 9/29/17. A customer reply is requested from all affected consignees. |
| Quantity | 44 units |
| Distribution | Worldwide distribution to US, Canada, France, and Germany. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K980367 | AEROSET SYSTEM Abbott Laboratories, Apr 1, 1998 |