Recall Z-0166-2018 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Abbott Laboratories, Inc

Tubing, Peristaltic Head Part Number 7-35009685-01 for use in clinical chemistry systems.

DateSep 29, 2017
StatusCompleted
ReasonThe Peristaltic Head tubing placed on the ARCHITECT c4000, c8000, and c16000 has the possibility to leak due to a manufacturing issue associated with the tubing connector.
Root causeComponent design/selection
ActionAll worldwide ARCHITECT Clinical Chemistry System customers with impacted instruments will receive the Product Correction Letter dated, 9/29/17. A customer reply is requested from all affected consignees.
Quantity44 units
DistributionWorldwide distribution to US, Canada, France, and Germany.