Recall Z-0047-2018 ยท KPS System, Tomography, Computed, Emission

Siemens Medical Solutions USA, Inc.

Torque wrench used on the bolts for the Symbia T2 model 10275008, Symbia Intevo Excel model 10764801, Symbia Intevo 16 model 10764804, Symbia Evo model 10910500, and Symbia Evo Excel model 10910501 devices.

DateSep 18, 2017
StatusTerminated
ReasonThe torque wrench used to tighten system bolts during installation was found to be out of tolerance
Root causeEquipment maintenance
ActionThe recall was initiated via phone and visit beginning 9/18/2017.
Quantity16 devices
DistributionDistribution was made to medical facilities located in CA, FL, GA, LA, NY, PA, and TN. There was government distribution but no foreign/military distribution.
FDA recordOpen on fda.gov