Recall Z-0044-2018 ยท KNW Instrument, Biopsy

Bard Peripheral Vascular Inc

BARD MAX-CORE Disposable Core Biopsy Instrument, Product Codes - MC1410, 14 g x 10 cm; MC1616, 16 g x 16 cm; MC1816, 18 g x 16 cm; MC1820, 18 g x 20 cm; MC1825, 18 g x 25 cm. The core needle biopsy device is intended for use in obtaining biopsies from soft tissues.

DateSep 22, 2017
StatusTerminated
ReasonDuring the review of May 2017 complaint data, a significant increase in MAX-CORE Disposable Core Biopsy Instrument complaints was observed for Failure to Prime, Failure to Fire, Failure to Obtain Sample, and Self-Activation.
Root causeDevice Design
ActionBPV sent a customer recall letter on September 22, 2017 to each of the US consignees via FedEx with proof of delivery notification. BPV will also initiate a recall for product outside the US in accordance with regulations of each country in which the product was distributed.
Quantity110,550 units
DistributionUS, Australia, Canada, Europe, Middle East, Africa, Latin America, Asia Pacific
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133948BARD(R) MONOPTY(R) DISPOSABLE CORE BIOPSY INSTRUMENT, BARD(R) MAX-CORE(R) DISPOSABLE CORE INSTRUMENT
Bard Peripheral Vascular, Inc., Feb 21, 2014