Recall Z-0027-2018 ยท NSU Instrumentation For Clinical Multiplex Test Systems

Hologic, Inc

Panther(R) System, Catalog Number 902615, For Use with Hologic Assays, For In Vitro Diagnostic Use

DateSep 15, 2017
StatusTerminated
ReasonAn incorrect incubator was installed on one Panther System which potentially led to incorrect Aptima Combo 2(R) assay (AC2) results.
Root causeEmployee error
ActionThe customer was notifed of the potential incorrect results via a phone call on September 15, 2017. A customer letter was sent on September 18, 2017. The notification letter requests the customer to re-test any negative Aptima Combo 2 assay results tested between July 13, 2017 Aug 30, 2017.
Quantity1 unit
DistributionIndiana
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K132251APTIMA COMBO 2 ASSAY (PANTHER SYSTEM)
Hologic / Gen-Probe Incorporated, Oct 17, 2013