Recall Z-0041-2018 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Encore Medical, Lp

EMPOWR PS KNEE(TM) e+, REF 343-13-708, SIZE 8, 13mm, QTY 01, SKY BLUE, STERILE Product Usage The EMPOWR PS Inserts are manufactured from ultra high molecular weight polyethylene (UHMWPE) infused with pure liquid pharmaceutical grade alpha-tocopheral and then cross-linked to conform to ASTM F2695. This insert is posterior stabilized and is intended to more closely match the kinematics of the k

DateSep 15, 2017
StatusTerminated
ReasonIt was reported that an implant labeled as 343-13-708 was packaged with implant 343-19-707.
Root causePackaging process control
ActionThe firm initiated the recall by email on 09/05/2017. The letter identified the affected product, problem and actions to be taken. The consignee was directed to notify anyone to whom the product was transferred and to return any in-stock product.
Quantity48 units
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov