ClearanceBrief
Recall Z-0057-2018 ยท LPB Cardiac Ablation Percutaneous Catheter

Biosense Webster, Inc.

THERMOCOOL SMARTTOUCH Uni-Directional Navigation Catheter (D133602S)

DateSep 6, 2017
StatusTerminated
ReasonBiosense Webster, Inc. has recently received an increased number of complaints related to the display of Alert 402 on the CARTO 3 System for certain lots of THERMOCOOL brand catheters. Alert 402 implies a "Map: magnetic distortion" when connected to CARTO 3 System. This issue may subsequently lead the physician to ablate in an unintended area when delivering RF energy.
Root causeUnder Investigation by firm
ActionAll US and OUS customers, who have purchased the lots identified as part of this recall will be notified via customer letters in order to return the product to BWI facility where it will be segregated for further disposition. This product recall applies to hospitals (or direct customers) and distributors who have purchased and/or received the THERMOCOOL Catheters identified as part of this recall . One notification package which includes the Customer Notification Letter, Business Reply Form (B RF) and pre-paid return label will be distributed to the customers subject to this recall via a 3rd
Quantity19 units
DistributionUS, Japan, Canada, Australia, New Zealand, Brazil, Croatia, Korea, Austria, Colombia, Italy, Portugal, Argentina, South Africa, Belgium, France, Germany, Ireland, Poland, Netherlands, Sweden, Denmark, Finland, Switzerland, Great Britain, Hungary, Czech Republic, China
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.